Pt 100 Prices

Pt100 prices can vary widely because “Pt100” describes the sensing principle, not the complete mechanical product. The same Pt100 temperature sensor can be supplied as a bare RTD element, a cable probe, a bayonet sensor, a threaded process probe, a terminal-head assembly or a complete transmitter-based temperature measurement system.
In simple terms: the price of a Pt100 is determined less by the word “Pt100” and more by how the sensor is built and where it must operate. Tolerance class, probe dimensions, wiring method, sheath material, cable type, process fitting, transmitter, calibration and quantity can all have a major effect on cost.
A Pt100 is a platinum resistance temperature detector (RTD) with a nominal resistance of 100 Ω at 0 °C. If you first want to understand the sensing principle, see What Is a Pt100 Sensor?.
Pt100 buying guide — quick navigation
What affects the price? · Element vs complete probe · Tolerance class · 2/3/4-wire connection · Probe and sheath · Transmitters · Calibration · Pt1000 / thermocouple alternatives · Information required for a quote
Why do Pt100 prices vary so much?
A single universal “Pt100 price” is not very meaningful because products sold under that description can have completely different constructions.
For example, all of the following may contain a Pt100 element:
bare platinum RTD element
small cable-mounted probe
stainless-steel industrial probe
spring-loaded bayonet sensor
threaded process temperature probe
terminal-head RTD assembly
thermowell-mounted temperature assembly
Pt100 with integrated 4–20 mA transmitter
The platinum sensing element may be similar, but the mechanical construction, protection level, cable system and electronics can be completely different.
Main factors that determine Pt100 cost
| Specification | Why it affects price |
| RTD tolerance class | Tighter tolerance can require selected elements and tighter production control |
| Thin-film / wire-wound element | Element construction and intended performance are different |
| 2-, 3- or 4-wire connection | Cable count, termination and measurement architecture change |
| Probe diameter and length | Material quantity, response time and manufacturing difficulty change |
| Sheath material | Standard stainless steel and special corrosion-resistant alloys have different costs |
| Cable type and length | PVC, silicone, PTFE and fiberglass have different temperature and environmental ratings |
| Process connection | Threads, compression fittings, flanges and bayonet fittings add mechanical components |
| Connection head | Adds terminals, enclosure protection and space for a transmitter |
| Temperature transmitter | Converts the RTD resistance into 4–20 mA or another process signal |
| Calibration | Calibration points, uncertainty and traceability add laboratory work |
| Order quantity | A one-off custom probe and a production batch do not have the same unit cost |
A bare Pt100 element and a complete Pt100 probe are not the same product
This is one of the biggest sources of confusion when comparing online prices.
A bare RTD is simply the platinum sensing element. An industrial probe may additionally contain:
metal sheath
electrical insulation
internal lead wires
external cable
process fitting
terminal block or connection head
For this reason, the price of a loose RTD chip should not be compared directly with the price of a complete industrial Pt100 probe.
Does Pt100 tolerance class affect the price?
It can. Platinum RTDs are available in different standardized tolerance classes.
Common temperature-tolerance expressions include:
Class AA: ±(0.10 + 0.0017 × |t|) °C
Class A: ±(0.15 + 0.002 × |t|) °C
Class B: ±(0.30 + 0.005 × |t|) °C
Class C: ±(0.60 + 0.010 × |t|) °C
A Pt100 accuracy class is therefore not one fixed ±°C value over every temperature. The permitted deviation changes with temperature, and the valid class range can also depend on element construction.
When comparing two Pt100 quotations, do not look only for “Class A.” Check over which temperature range that class is actually specified.
Does a more expensive Pt100 automatically mean a more accurate Pt100?
No. Price and temperature accuracy are not the same thing.
A high-priced probe may cost more because it has:
a long stainless-steel sheath · special process fitting · high-temperature cable · weatherproof terminal head · transmitter · custom fabrication
rather than because its bare RTD element has tighter tolerance.
A mechanically simple Class A probe can therefore be a better measurement choice than a much more expensive assembly that was designed for an entirely different process.
Thin-film vs wire-wound Pt100: does construction affect cost?
Pt100 elements are commonly manufactured as thin-film or wire-wound sensing resistors.
Thin-film Pt100: a platinum layer is deposited onto a ceramic substrate. Compact dimensions, low thermal mass and efficient mass production are major advantages.
Wire-wound Pt100: fine platinum wire is supported or wound in an insulating structure. Depending on the design, it can be attractive for particular precision or wider-temperature applications.
Neither construction should be selected from price alone. Operating temperature, vibration, response time and required tolerance are more important.
Why do 2-wire, 3-wire and 4-wire Pt100 probes have different prices?
Because Pt100 measures resistance, the resistance of the connecting wires can become part of the measurement.
2-wire Pt100: simplest connection. Both lead resistances add directly to the RTD resistance.
3-wire Pt100: suitable RTD electronics compensate most of the lead resistance when the relevant conductors are well matched.
4-wire Pt100: separates current and voltage measurement paths using a Kelvin connection, giving the best rejection of lead resistance.
The extra wires therefore do more than increase cable cost; they change the measurement method.
Pt100 or Pt1000 for a long cable run?
For long 2-wire installations, a Pt1000 can also be worth considering.
Pt1000 has approximately ten times the nominal resistance and approximately ten times the Ω/°C change of Pt100, so the same copper lead resistance creates a smaller equivalent temperature error.
Another solution is to place a transmitter close to the Pt100 and send the long-distance signal as 4–20 mA.
Does probe diameter affect Pt100 price?
Yes. Probe diameter affects both manufacturing and process behaviour.
A small-diameter probe can provide faster thermal response but may require more delicate construction and can have lower mechanical strength.
A larger probe can be mechanically stronger, but additional thermal mass can slow its response.
Two probes containing the same Pt100 element can therefore behave very differently in the process.
Does Pt100 probe length affect cost?
A longer probe uses more sheath and internal material, but the more important issue is correct immersion depth.
If the sensor is too short, heat conducted through the process fitting or surrounding air can influence the sensing region.
Reducing price by choosing an unnecessarily short probe can therefore create a larger measurement error than the sensor tolerance itself.
How does sheath material affect price?
Stainless steel is common in industrial Pt100 probes, but chemical, food, high-temperature and corrosive processes may require different materials.
The sheath must be suitable for:
process temperature
chemical exposure
pressure
vibration
mechanical impact
A cheaper sheath that fails prematurely in the process is rarely the lowest-cost solution over the life of the machine.
Why does Pt100 cable type affect the price?
Cable length matters, but insulation material can matter even more.
Common options include:
PVC
silicone
PTFE
fiberglass
braided or mechanically protected cable
A high-temperature cable naturally costs more than basic PVC, but using unsuitable cable next to a hot process can result in premature failure.
Why does a bayonet Pt100 cost differently?
A bayonet probe includes mechanical components such as a spring, bayonet cap and adjustable contact arrangement in addition to the RTD element.
For plastics machinery, moulds and similar applications, the dimensions of the spring-loaded assembly and probe must match the installation. See the bayonet-type temperature sensor section for this mechanical family.
Do process fittings affect Pt100 cost?
Yes. Fixed threads, adjustable compression fittings, flanges and custom mounting hardware all add material and machining.
More importantly, process connection is a safety issue. A probe fitted to a pressurized vessel or pipeline must have a connection appropriate to the mechanical process conditions.
Is a temperature transmitter included in the Pt100 price?
Not necessarily. A Pt100 itself is a passive resistance sensor.
A temperature transmitter measures the RTD resistance, calculates temperature and can convert the result into:
4–20 mA
0–10 V
digital communication
A “Pt100 + transmitter” assembly is therefore not equivalent to a simple Pt100 probe and should not be compared at the same price level.
When is a Pt100 transmitter useful?
A transmitter is particularly useful when:
the cable run is long
the electrical environment is noisy
the PLC does not have an RTD input
a standard process signal is required
Measuring the resistance close to the sensor and transmitting 4–20 mA over the longer cable can provide a more robust industrial signal path.
Can Pt100 connect directly to a PLC?
Yes, when the PLC has a compatible RTD input.
Look for Pt100, IEC 60751 or 2/3/4-wire RTD support in the input-module specification.
A Pt100 should not be connected to an ordinary 0–10 V analog input as though it were a voltage-output sensor. If no RTD input is available, use an appropriate transmitter.
Can Pt100 connect directly to a temperature controller?
Many temperature controllers support Pt100 directly.
The input type must still be configured correctly. A controller left in Type K thermocouple mode will not interpret a Pt100 resistance correctly.
Once the Pt100 measurement is available, the controller can operate heating equipment through outputs such as an SSR, contactor or another power-control device.
Does calibration affect Pt100 price?
Yes. A standard production sensor and an individually calibrated, documented sensor are not the same delivery.
Calibration cost can depend on:
number of calibration points
required uncertainty
traceability requirements
certificate scope
There is also no universal rule that every Pt100 must be recalibrated at one fixed interval. Calibration frequency should be based on process criticality, historical drift, quality-system requirements and the uncertainty required from the measurement.
Is calibration the same as Pt100 tolerance class?
No.
A Class A or Class B designation describes the tolerance that the sensor is manufactured to meet under defined conditions.
Calibration measures the actual behaviour of an individual sensor against a reference at specified temperature points.
The two pieces of information are related to measurement quality but are not interchangeable.
Pt100 or Pt1000: which is more economical?
The sensor-element price alone does not answer this question.
Pt1000 can reduce the relative influence of lead resistance in long 2-wire installations because its nominal resistance is approximately ten times higher.
Pt100, however, is extremely common in industrial instrumentation and has broad support across probes, transmitters, controllers and PLC modules.
For a fair comparison, evaluate the complete system cost: sensor + cable + RTD input + transmitter if required.
Is a thermocouple cheaper than Pt100?
There is no universal answer.
A simple Type J or Type K thermocouple can be very economical. Noble-metal thermocouples such as R, S or B can be considerably more expensive.
Pt100 versus thermocouple selection should therefore consider temperature range, atmosphere, accuracy, response time, wiring and mechanical construction as well as purchase price.
For a full engineering comparison, see Thermocouple vs Pt100.
If PTC or NTC is cheaper, why use Pt100?
PTC and NTC thermistors are excellent sensors for many applications, but they are not electrically equivalent to a Pt100 RTD.
Pt100 offers a standardized platinum R/T characteristic, broad industrial instrument support and excellent measurement stability when correctly applied.
For motor over-temperature protection, however, a switching PTC thermistor may be a much better choice than an RTD if the requirement is simply to detect that a thermal limit has been exceeded.
Can a non-contact sensor replace Pt100?
Where the target is moving, inaccessible or too hot for direct contact, non-contact temperature sensors can be considered.
Infrared measurement has different error sources, including emissivity, field of view and optical path. Pt100 measures by direct thermal contact.
These technologies should therefore not be selected by sensor price alone.
Why don't we publish a fixed Pt100 price range?
Static statements such as “low-cost Pt100 sensors are 20–50 USD and high-quality units are 100–500 USD” quickly become misleading.
They can also accidentally compare completely different products, such as:
a short 3 mm probe with a one-metre process assembly,
a standard Class B sensor with an individually calibrated precision system,
a bare RTD element with a transmitter-equipped terminal-head assembly.
A better approach is to define the actual required construction and then compare current prices for that specification.
What information should you provide when requesting a Pt100 price?
For an accurate quotation, specify as much of the following as possible:
1. Minimum and maximum process temperature
2. Required Pt100 tolerance class
3. 2-, 3- or 4-wire connection
4. Probe diameter
5. Probe length / insertion length
6. Sheath material
7. Cable length
8. Cable insulation type
9. Process connection
10. Bayonet, threaded, flanged or straight probe construction
11. PLC or controller interface
12. Whether a transmitter is required
13. Calibration certificate requirement
14. Quantity
Without these details, two quotations labelled “Pt100” may not represent technically comparable products at all.
Brand or specification: which matters more?
A reliable manufacturer is valuable, but brand name alone is not enough.
Before purchasing, verify:
platinum characteristic
tolerance class
probe construction
operating temperature
cable type
wiring method
process compatibility
A cheap sensor that is wrong for the process can cost far more over the life of the machine than a correctly specified probe with a higher initial purchase price.
Conclusion: how should Pt100 price be evaluated?
Pt100 price should never be judged from the sensor name alone.
The real cost is determined by the complete specification: RTD element, tolerance class, 2/3/4-wire construction, probe dimensions, sheath, cable, process fitting, transmitter, calibration and quantity.
The most expensive Pt100 is not automatically the best sensor for every application. Correct process compatibility is more important than the price label.
Once correctly selected, Pt100 can form the first stage of a complete industrial measurement chain extending through a temperature transmitter, PLC, HMI and SCADA system.
Pt100 buying and temperature-measurement knowledge map
Pt100 / Pt1000 and the platinum RTD family
Pt100 / Pt1000 Main Category · What Is Pt100? · What Is Pt1000? · Bayonet-Type Temperature Sensors
Pt100 or thermocouple?
Thermocouple vs Pt100 · What Is a Thermocouple? · Thermocouple Types · Type K Thermocouples · Type J Thermocouples
PTC, NTC and other resistance-based sensors
PTC / NTC Sensors · PTC Sensors · What Is PTC? · NTC Sensors · What Is NTC?
Non-contact temperature measurement