What Is a Type S Thermocouple?

A Type S thermocouple is a noble-metal thermocouple made from a platinum-10% rhodium positive thermoelement and a platinum negative thermoelement. It is widely associated with high-temperature measurement, calibration systems and temperature metrology.

In simple terms: a Type S thermocouple produces a very small thermoelectric voltage that changes with temperature. A temperature controller, PLC or precision measuring instrument interprets that voltage using the standardized Type S temperature–EMF reference function.

Type S should not be thought of simply as an expensive Type K thermocouple for higher temperatures. Its platinum-based thermoelements, relatively low EMF sensitivity, susceptibility to contamination and frequent use with high-purity ceramic protection place it in a very different application class.

What materials are used in a Type S thermocouple?

The nominal Type S thermoelement pair is:

Positive (+): Platinum-10% Rhodium, Pt-10%Rh
Negative (-): Platinum, Pt

Because both thermoelements are based on platinum-group precious metals, Type S belongs to the noble-metal thermocouple family. Type R and Type B thermocouples belong to the same broader group.

What is the Type S reference temperature range?

The ITS-90 reference function for Type S is defined over approximately -50 °C to +1768 °C.

This does not mean that every physical Type S probe can operate continuously throughout that entire range.

The reference-function range describes the standardized mathematical temperature–EMF relationship. The real service limit of a sensor also depends on:

thermoelement diameter · ceramic insulation · protection tube · process atmosphere · contamination · mechanical construction · continuous or intermittent exposure.

This distinction is explained further in IEC 60584-1 Thermocouple EMF Reference Functions and Tolerances.

Why is Type S used at high temperatures?

Platinum and platinum-rhodium alloys can provide useful chemical stability and repeatable thermoelectric behaviour at elevated temperatures when they are correctly protected and used in a suitable atmosphere.

For this reason, Type S can be found in high-temperature furnaces, ceramic processing, calibration equipment, heat-treatment research and specialized industrial temperature measurement.

High-temperature capability does not mean immunity to environmental effects. Contamination is one of the most important causes of drift in platinum-based thermocouples.

Why are Type S thermocouples expensive?

Both thermoelements contain precious metals. The positive leg contains platinum and rhodium, while the negative leg is platinum.

Raw-material cost is therefore much higher than for base-metal thermocouples such as J, K, T, E and N.

The final sensor price can also depend on:

wire diameter · thermoelement length · platinum/rhodium value · ceramic insulation · protection tube · calibration · terminal head · custom construction.

For this reason, a permanent fixed-price table is rarely meaningful. See Thermocouple Prices for the main cost factors.

How does a Type S thermocouple work?

Like other thermocouples, Type S operates through the Seebeck effect.

The Pt-10%Rh and Pt thermoelements have different thermoelectric properties. When they form a circuit exposed to a temperature distribution, a small electromotive force is produced.

That voltage is not a temperature value by itself. The measuring instrument must apply the standardized Type S reference function to convert EMF into temperature.

Why is the Type S output signal relatively small?

Type S has lower thermoelectric sensitivity than many common base-metal thermocouples.

Even at high temperatures, its output remains in the millivolt range. This places greater demands on the measurement electronics.

A good Type S input should provide:

low offset
low electrical noise
stable gain
accurate cold-junction compensation

In precision calibration work, even a few microvolts of unwanted error can become significant.

Does Type S require cold-junction compensation?

Yes. Using platinum thermoelements does not eliminate the need for a reference-junction correction.

Type S reference functions assume a reference junction at 0 °C. A real PLC, transmitter or measuring instrument normally operates with its terminals at another temperature, so it must apply cold-junction compensation (CJC).

Because Type S produces relatively low EMF per degree, reference-junction errors can become an important part of the overall measurement uncertainty.

Type S temperature–EMF reference function

The forward reference relationship can be written conceptually as:

E = f(t90)

where E is thermocouple EMF and t90 is temperature on the ITS-90 scale.

A measuring instrument performs the inverse operation:

t90 = g(E)

The relationship between reference tables, polynomial functions and ITS-90 is covered in Thermocouple Reference Tables, Polynomials and ITS-90.

Where are Type S thermocouples used?

With an appropriate protection system, Type S can be used in applications such as:

High-temperature furnaces
Ceramic and heat-treatment processes
Calibration laboratories
High-temperature test equipment
Metallurgical research
Reference and transfer thermometry

Whether the thermoelements can be exposed directly to the process is a separate question. Protection-tube material, ceramic insulation and furnace atmosphere can be just as important as the thermocouple type itself.

What atmosphere is suitable for Type S?

Platinum and platinum-rhodium thermocouples generally perform best in clean oxidizing or neutral atmospheres.

Strongly reducing atmospheres, metal vapours and unsuitable refractory materials can contaminate platinum thermoelements at high temperature.

Elements such as silicon and various metallic contaminants can diffuse into platinum and change its local thermoelectric properties.

For this reason, a high-temperature Type S application should ask not only “How hot will it be?” but also “What atmosphere and protection materials will surround the thermocouple?”

Why is high-purity ceramic protection commonly used?

High-temperature Type S assemblies often use high-purity ceramic insulators and protection tubes to keep the thermoelements separated and reduce contamination.

High-purity alumina, for example, can be useful in suitable applications because it helps isolate platinum thermoelements from foreign metals and process contaminants.

No protection system is unlimited, however. During long exposure at very high temperature, impurities from surrounding materials can still migrate toward the thermoelements.

Can a metal protection tube be used with Type S?

It can be used in certain applications, but the complete temperature and chemical environment must be considered carefully.

At high temperatures, direct or indirect interaction between platinum thermoelements and unsuitable metallic materials can cause contamination and drift.

For demanding high-temperature Type S measurement, ceramic protection systems are therefore common.

Why is contamination so important with Type S?

Accurate thermocouple behaviour depends on the composition and homogeneity of the thermoelements.

If foreign elements diffuse into platinum or platinum-rhodium wire at high temperature, the local Seebeck coefficient of the wire can change.

The thermocouple can still have electrical continuity and appear healthy on a multimeter while its calibration has shifted.

A thermocouple that is not open-circuit is not necessarily an accurate thermocouple.

Can a Type S thermocouple drift with time?

Yes. Like any real thermocouple used for long periods at high temperature, Type S can experience thermoelectric drift.

Possible causes include:

contamination · local changes in alloy composition · mechanical strain · loss of thermoelectric homogeneity · prolonged high-temperature exposure.

Critical processes may therefore require periodic calibration, comparison against another sensor or planned replacement intervals.

Was Type S historically used as a temperature standard?

Yes. Type S has an important place in the history of temperature metrology.

Under the former IPTS-68 temperature scale, the Pt-10%Rh/Pt thermocouple was used as a standard interpolating instrument over part of the high-temperature region.

With the introduction of ITS-90, Type S no longer serves as a defining standard interpolation thermometer for the international temperature scale.

It nevertheless remains important in high-temperature calibration and as a reference or transfer thermocouple.

Type S versus Type R

Types S and R are closely related noble-metal thermocouples:

Type S: Pt-10%Rh / Pt
Type R: Pt-13%Rh / Pt

Their temperature–EMF reference functions are different. A Type S thermocouple must therefore not be connected to an instrument configured for Type R and expected to read correctly.

See What Is a Type R Thermocouple? for the detailed comparison.

Type S versus Type B

Type B uses Pt-30%Rh against Pt-6%Rh and is particularly associated with very high-temperature measurement.

Unlike Type S, where the negative leg is platinum, both Type B thermoelements are platinum-rhodium alloys.

Its reference function and high-temperature behaviour are different, so Type B is not simply a direct replacement for Type S when a higher temperature is required.

See What Is a Type B Thermocouple?.

Type S versus Type K

Type K is a nickel-based base-metal thermocouple used extensively in general industrial machinery.

Type S is a platinum-rhodium/platinum noble-metal thermocouple used where higher temperatures, specialized furnace conditions or higher-level calibration requirements justify its cost and protection requirements.

A Type S sensor is not automatically more accurate than Type K at every temperature simply because it is more expensive. Sensor choice must match the actual process.

Type S versus Pt100

Type S measures temperature through thermoelectric EMF, while a Pt100 measures the change in electrical resistance of a platinum sensing element.

Pt100 sensors can provide excellent stability and accuracy over many low- and moderate-temperature applications. Type S extends into much higher-temperature regions.

For a broader comparison, see Thermocouple vs Pt100.

Can Type S be connected directly to a PLC?

Yes, provided the PLC thermocouple module specifically supports Type S.

The PLC thermocouple input must be configured for Type S. The module measures the low-level EMF, measures its reference-junction temperature, applies CJC and converts the compensated signal using the Type S reference function.

Because Type S produces a very small signal, a suitable low-level thermocouple input is required. Connecting it directly to a standard 0–10 V analog input is not an appropriate measurement method.

Can Type S be converted to 4–20 mA?

Yes. A compatible temperature transmitter or sensor converter can measure the Type S thermocouple and convert the calculated temperature into a standard process signal.

The transmitter performs Type S linearization and cold-junction compensation and can output a signal such as 4–20 mA.

This can be useful when the distance between the temperature sensor and the control system is large.

Can extension or compensating cable be used with Type S?

Yes, but the correct cable system must be selected.

Running platinum and platinum-rhodium conductors all the way from the furnace to the control room can be unnecessarily expensive.

Suitable compensating cables can therefore be used over specified temperature ranges. These conductors are not made from the same noble-metal alloys as the thermocouple; they are designed to reproduce an appropriate thermoelectric relationship over their defined operating range.

Cable requirements are discussed under IEC 60584-3 Thermocouple Extension and Compensating Cables.

Can Type S be grounded or ungrounded?

Different mechanical and electrical constructions are possible, but the usual high-temperature Type S installation is often different from a conventional stainless-steel mineral-insulated Type K probe.

Platinum thermoelements are frequently supported by ceramic insulators and installed inside suitable ceramic protection tubes.

For the general meaning of junction construction, see Grounded, Ungrounded and Exposed-Junction Thermocouple Differences.

What does Type S tolerance mean?

A real Type S thermocouple can deviate slightly from the ideal standardized reference function. The permitted deviation is defined by the applicable thermocouple tolerance.

Tolerance is not the same as individual calibration. Two Type S thermocouples can satisfy the same tolerance requirements and still produce slightly different actual EMF values at the same temperature.

For current IEC terminology and tolerance interpretation, see Thermocouple Standards.

Why is calibration important for Type S?

Type S is frequently used in applications where high-temperature measurement quality matters, so individual calibration can be particularly valuable.

A Type S thermocouple can be calibrated at suitable fixed points or by comparison against a reference temperature system.

Calibration determines the actual deviation of one specific thermocouple from the standardized Type S reference function. Measurement can then be based on sensor-specific correction data rather than nominal tolerance alone.

Why is total uncertainty more than Type S tolerance?

In a high-quality Type S measurement system, uncertainty can include contributions from:

thermocouple calibration · thermoelement drift · contamination · compensating cable · cold-junction compensation · measurement electronics · furnace gradients · immersion depth · temperature uniformity.

A single tolerance figure therefore does not describe the performance of the complete measurement chain.

The complete approach is described in Measurement Uncertainty in a Thermocouple System.

What should be specified when selecting a Type S thermocouple?

A useful Type S specification should consider at least:

Normal process temperature
Maximum temperature
Continuous or intermittent exposure
Oxidizing, neutral or reducing atmosphere
Thermoelement diameter
Ceramic insulation type
Protection-tube material
Insertion length
Terminal-head arrangement
Type S support in the PLC or controller
Calibration requirement and target uncertainty

At very high temperatures, the compatibility of the process atmosphere, ceramic system and thermoelements can be more important to service life than the thermocouple letter alone.

Advantages of Type S thermocouples

In the correct application, Type S can provide:

High-temperature measurement capability
Well-characterized platinum-based thermoelectric behaviour
Suitability for high-level calibration and metrology applications
Long-established standardized reference functions

Limitations of Type S thermocouples

Important limitations include:

High raw-material cost
Lower EMF sensitivity than many base-metal thermocouples
Susceptibility to contamination at high temperature
Need for suitable ceramic insulation and protection
Requirement for cold-junction compensation
Possible long-term drift

Conclusion

Type S is a Pt-10%Rh / Pt noble-metal thermocouple with an important role in high-temperature measurement and precision thermometry.

Its ITS-90 reference function covers approximately -50 °C to +1768 °C, but that mathematical range must not be confused with the guaranteed service range of every physical Type S sensor.

Real Type S performance depends heavily on platinum purity, thermoelement homogeneity, contamination, process atmosphere, ceramic protection and calibration.

A reliable Type S system therefore requires more than an expensive sensor. Correct protection, suitable compensating cable, accurate cold-junction compensation, Type S reference linearization and, where required, individual calibration must all be considered together.

Related thermocouple types and temperature measurement topics

+ Types of Thermocouples (T/C)

+ What Is a Type R Thermocouple? · What Is a Type B Thermocouple?

+ What Is a Type T Thermocouple? · Type K Thermocouples

+ What Is a Thermocouple (T/C)? · Seebeck Effect

+ Thermocouple Cold-Junction Compensation

+ Thermocouple Standards · IEC 60584-1

+ IEC 60584-3 Extension and Compensating Cables

+ Pt100 Sensors · Thermocouple vs Pt100

+ Temperature Sensor Converters · Temperature Controllers · PLC Systems