What Is a Type B Thermocouple?
A Type B thermocouple is a noble-metal thermocouple made from a platinum-30% rhodium (Pt-30%Rh) positive thermoelement and a platinum-6% rhodium (Pt-6%Rh) negative thermoelement. It is intended primarily for high- and very-high-temperature measurement and belongs to the platinum-rhodium family together with Type R and Type S thermocouples.
In simple terms: Type B is a thermocouple designed mainly for very hot furnaces and similar processes. Unlike Types R and S, both Type B thermoelements contain rhodium.
Type B also has an unusual electrical characteristic: its thermoelectric sensitivity is extremely small near room temperature but becomes much more useful at high temperature. For this reason, simply describing Type B as a “0 to 1820 °C thermocouple” misses much of what makes it different.
What materials are used in a Type B thermocouple?
The nominal thermoelement pair is:
Positive (+): Platinum-30% Rhodium — Pt-30%Rh
Negative (-): Platinum-6% Rhodium — Pt-6%Rh
This construction makes Type B fundamentally different from nickel-based Type K, Type N and Type E thermocouples.
It also differs from its platinum relatives:
Type R: Pt-13%Rh / Pt
Type S: Pt-10%Rh / Pt
Type B: Pt-30%Rh / Pt-6%Rh
In Type B, even the negative thermoelement contains rhodium, giving the thermocouple a distinctly different high-temperature temperature–EMF characteristic.
What is the Type B reference temperature range?
The ITS-90 reference function for Type B is defined over approximately 0 °C to +1820 °C.
That does not mean every physical Type B probe is rated for continuous service from 0 to 1820 °C.
The reference-function range and the practical operating range of a real sensor are not the same thing.
Actual service capability depends on:
thermoelement diameter
platinum-rhodium purity
ceramic insulation
protection tube
process atmosphere
contamination
exposure time
The distinction between standardized thermoelectric data and real probe limits is explained further in IEC 60584-1 Thermocouple EMF Reference Functions and Tolerances.
At what temperatures is Type B actually useful?
Although the reference function begins at 0 °C, Type B is primarily a high-temperature thermocouple.
Its practical value becomes much greater in furnace-temperature regions where platinum-rhodium construction provides useful stability and mechanical performance.
At temperatures of only 100, 200 or 300 °C, Type B normally provides little practical advantage. A Pt100, Type J, K, T, E or N will often be more appropriate depending on the application.
The reason is not that Type B suddenly stops functioning at low temperature. The real issue is its very low thermoelectric sensitivity in that region.
Why does Type B produce such a small signal near room temperature?
Both thermoelements are platinum-rhodium alloys, and their thermoelectric behaviour is relatively similar near room temperature.
When the Seebeck coefficients of the two conductors approach one another, their difference becomes small, so the net thermocouple EMF changes only very slightly with temperature.
The physical principle behind this behaviour is the Seebeck effect.
This low sensitivity also means that electrical noise and low-level instrument errors can represent a much larger fraction of the useful Type B signal at low temperatures.
Why is the Type B EMF curve unusual near room temperature?
The Type B reference curve behaves differently from the simple mental picture of “temperature rises, voltage always rises at a fixed rate.”
Close to room temperature, Type B produces only a few microvolts of change and its reference EMF passes through a very shallow region.
This unusual low-temperature behaviour is one reason Type B is rarely selected for ordinary ambient or moderate-temperature process measurements.
Its real strength lies much farther up the temperature scale.
How does a Type B thermocouple work?
Like the other thermocouple types, Type B produces an EMF because its two thermoelements respond differently to a temperature distribution.
The Seebeck coefficients of Pt-30%Rh and Pt-6%Rh change with temperature, so the resulting EMF is also temperature dependent and nonlinear.
A measuring instrument must therefore use the actual Type B reference function rather than a fixed millivolts-per-degree conversion.
Type B temperature–EMF reference function
The forward relationship can be represented conceptually as:
E = f(t90)
A measuring instrument performs the inverse calculation:
t90 = g(E)
Type B is included among the letter-designated thermocouples covered by IEC 60584-1, with reference functions based on the ITS-90 temperature scale and a 0 °C reference junction.
For the mathematics behind the tables and polynomials, see Thermocouple Reference Tables, Polynomials and ITS-90.
Why is Type B different from Type R and Type S at high temperature?
The negative thermoelement of Types R and S is essentially platinum. In Type B, the negative thermoelement is Pt-6%Rh.
This all-alloy construction gives Type B different mechanical and thermoelectric behaviour at elevated temperature.
In suitable applications, Type B can provide:
a higher useful temperature region
good mechanical strength at elevated temperature
useful stability in carefully controlled high-temperature environments
This is why Type B is normally considered for a different part of the temperature spectrum than Types R and S.
Type B versus Type S
Type S: Pt-10%Rh / Pt
Type B: Pt-30%Rh / Pt-6%Rh
Type S uses pure platinum for the negative leg; Type B contains rhodium in both thermoelements.
Type S provides considerably more useful thermoelectric sensitivity at lower temperatures, while Type B is more strongly associated with the highest-temperature platinum-thermocouple applications.
Type B is therefore not simply a “better Type S.” They serve different operating regions.
Type B versus Type R
Type R: Pt-13%Rh / Pt
Type B: Pt-30%Rh / Pt-6%Rh
Like Type S, Type R uses platinum for its negative thermoelement.
Type R generally provides a stronger useful signal in lower-temperature regions, while Type B extends into a higher temperature region.
A Type B thermocouple must never be interpreted using Type R linearization.
Type B versus Type C
Type C uses W-5%Re / W-26%Re tungsten-rhenium thermoelements and can extend into much higher temperature regions than Type B.
However, Type C has a major restriction: tungsten-rhenium thermoelements are unsuitable for direct high-temperature exposure to oxidizing atmospheres.
Type B, by contrast, can be used in suitable clean oxidizing or inert high-temperature environments.
This means a Type B sensor can be the more appropriate choice in a 1600–1700 °C oxidizing furnace even though Type C has the higher theoretical temperature capability.
What atmospheres are suitable for Type B?
Type B thermocouples are commonly used in clean oxidizing or inert atmospheres at high temperature.
Platinum-rhodium thermoelements can also be used in some carefully controlled neutral or vacuum systems, depending on the complete sensor construction.
Potentially damaging environments include:
strongly reducing atmospheres
metal vapours
silicon-bearing contaminants
reactive process gases capable of contaminating platinum-group metals
At high temperature, even small amounts of contamination can alter the local composition of a thermoelement and change its thermoelectric behaviour.
Why is high-purity ceramic protection used with Type B?
The Pt-Rh thermoelements should be protected from metallic contamination and unwanted chemical interaction.
High-purity alumina and other suitable ceramic materials are therefore commonly used as electrical insulation and protection in high-temperature Type B assemblies.
The ceramic does more than keep the two wires electrically separated. It also forms part of the contamination-control system surrounding the platinum-rhodium thermoelements.
At these temperatures, sensor design must consider thermoelements, ceramic purity, protection tube and process atmosphere as one system.
Why should Type B thermoelements be protected from unsuitable metals?
At elevated temperatures, metallic vapours or direct contact with unsuitable alloys can introduce foreign elements into platinum-rhodium conductors.
That contamination can change thermoelement composition and therefore alter the local Seebeck coefficient.
For demanding high-temperature work, the thermoelements are therefore often isolated from metallic protection materials by suitable high-purity ceramics.
What happens when a Type B thermocouple becomes contaminated?
A thermocouple does not have to become open-circuit before it becomes inaccurate.
Foreign elements entering the platinum-rhodium conductors can change their local thermoelectric properties.
The sensor may still show normal continuity on a multimeter while its temperature–EMF relationship has shifted away from its original calibration.
Electrical continuity is not proof of thermocouple accuracy.
Can rhodium oxidation affect Type B?
Yes. Platinum-rhodium thermoelements can undergo compositional changes during prolonged high-temperature exposure.
Preferential oxidation or redistribution involving rhodium can change the effective alloy composition along portions of the conductor.
This can create thermoelectric inhomogeneity and contribute to calibration drift.
Material purity, clean ceramic protection and appropriate atmosphere are therefore important parts of a high-quality Type B installation.
Can a Type B thermocouple drift over time?
Yes. Noble-metal construction does not mean the sensor remains unchanged indefinitely.
Long-term high-temperature drift can result from:
contamination
changes in rhodium distribution
loss of thermoelectric homogeneity
mechanical strain
prolonged thermal ageing
Critical furnace and laboratory applications may therefore require periodic comparison, calibration or planned sensor replacement.
Does Type B require cold-junction compensation?
Type B is unusual in this respect, but the underlying thermocouple physics does not change.
Its standardized reference function is defined relative to a 0 °C reference junction, so the principle of cold-junction compensation (CJC) still applies.
However, the Type B EMF changes only slightly over temperatures around normal room conditions. This means a reference-junction temperature variation in that region can produce a much smaller thermoelectric error than it would for many other thermocouple types.
This should not be interpreted as “Type B never needs CJC.” If the measuring instrument provides correct Type B cold-junction compensation, there is generally no reason to disable it.
Can Type B use compensating cable?
Yes. Running expensive platinum-rhodium conductors all the way from a furnace to a distant control panel is often unnecessary.
Suitable compensating cable can be used over its specified temperature region.
The compensating conductors do not necessarily use the same Pt-30%Rh and Pt-6%Rh alloys as the actual thermocouple. Instead, they are designed to provide a suitable thermoelectric approximation over a defined temperature range.
The general rules for thermocouple wiring are covered in IEC 60584-3 Thermocouple Extension and Compensating Cables.
Can Type B be connected directly to a PLC?
Yes, if the PLC thermocouple input specifically supports Type B.
A suitable PLC thermocouple module measures the low-level EMF, accounts for its terminal temperature and applies the Type B temperature–EMF function.
A Type B thermocouple should not be connected directly to an ordinary 0–10 V analog input. Its signal is only at millivolt level and requires appropriate low-level measurement and linearization.
Can Type B be converted to 4–20 mA?
Yes. A Type B-compatible temperature transmitter can measure the thermocouple locally.
The transmitter measures the low-level EMF, performs the required compensation and Type B linearization, then converts the temperature into a standard signal such as 4–20 mA.
The control system can then receive the result through a standard analog input.
Can Type B be connected to a temperature controller?
A temperature controller that explicitly supports Type B can accept the thermocouple directly.
The input must be configured as Type B. Selecting Type R or Type S causes the controller to use a different temperature–EMF function and produces an incorrect temperature.
At furnace temperatures around 1500–1700 °C, even relatively small measurement errors can become important to product quality and process repeatability.
Can non-contact measurement replace Type B?
At very high temperatures, inserting a physical sensor into the process can itself create materials and maintenance problems.
Some furnace applications therefore use non-contact temperature sensors, infrared pyrometers or optical temperature systems either alongside or instead of Type B.
Non-contact measurement has its own uncertainty sources, including emissivity, optical access, window transmission, field of view and surface condition.
It is therefore a different measurement method rather than an automatically superior one.
Type B versus Pt100
Type B measures temperature through thermoelectric EMF, while a Pt100 measures the resistance change of a platinum element.
Pt100 is highly useful for many low- and moderate-temperature measurements. Type B occupies almost the opposite end of the temperature spectrum and is designed for very hot processes.
At around 100 °C a Pt100 will usually be a far more sensible choice than Type B; at around 1600 °C, a Pt100 is no longer a suitable sensor technology.
See Thermocouple vs Pt100 for the general comparison.
What does Type B thermocouple tolerance mean?
A real Type B thermocouple can differ slightly from the ideal standardized reference function. The permitted deviation is described by its thermocouple tolerance.
That tolerance is not the accuracy of the complete installed furnace-measurement system.
Additional contributors can include:
thermoelement tolerance or calibration
high-temperature drift
contamination
compensating cable
cold-junction compensation
measurement electronics
furnace temperature gradients
sensor location and immersion
The distinction between tolerance and total measurement performance is discussed further under Thermocouple Standards.
Can a Type B thermocouple be calibrated?
Yes. Type B is an important noble-metal thermocouple for high-temperature calibration work.
Calibration determines the actual deviation of one physical thermocouple from the ideal Type B reference function at selected temperatures.
For critical furnace or laboratory measurements, individual calibration data can therefore be more informative than relying only on nominal thermocouple tolerance.
What contributes to Type B measurement uncertainty?
The uncertainty of a Type B system can include contributions from:
calibration
thermoelectric drift
rhodium-composition changes
contamination
compensating cable
CJC
measurement electronics
furnace uniformity
radiation and conduction effects
sensor position
Critical measurements should therefore consider the complete thermocouple measurement uncertainty rather than relying on a single ±°C value.
Where are Type B thermocouples used?
With the correct protection system, Type B can be used in:
high-temperature industrial furnaces
glass production
ceramic processing
metallurgical processes
high-temperature laboratories
heat-treatment and sintering systems
high-temperature calibration work
The thermocouple letter alone does not determine whether a probe is suitable. Atmosphere, ceramic protection, thermoelement diameter and the complete mechanical assembly must also match the process.
What should be specified when selecting a Type B thermocouple?
A useful specification should include:
Normal process temperature
Maximum temperature
Continuous or intermittent service
Oxidizing, inert or other process atmosphere
Thermoelement wire diameter
Ceramic insulation material
Protection-tube construction
Probe length and immersion depth
Type B support in the PLC or controller
Calibration requirement and target uncertainty
Advantages of Type B thermocouples
In the right application, Type B can provide:
very-high-temperature measurement capability
a higher useful upper-temperature region than Types R and S
good high-temperature mechanical properties
use in suitable clean oxidizing or inert atmospheres
suitability for high-temperature calibration and laboratory work
Limitations of Type B thermocouples
Important limitations include:
high platinum and rhodium cost
very low thermoelectric sensitivity at low temperatures
susceptibility to contamination at high temperature
need for suitable high-purity ceramic protection
limitations in reducing or chemically aggressive atmospheres
possible long-term thermoelectric drift
Conclusion
Type B is a Pt-30%Rh / Pt-6%Rh noble-metal thermocouple intended primarily for high- and very-high-temperature measurement.
Its ITS-90 reference function covers approximately 0 °C to +1820 °C, but that mathematical range should not be interpreted as the continuous-service rating of every Type B sensor.
The feature that most clearly separates Type B from Types R and S is that both thermoelements are platinum-rhodium alloys.
Its very small low-temperature EMF makes it a poor choice for ordinary moderate-temperature measurement, while its high-temperature construction can make it extremely valuable in furnaces, laboratories and specialized thermal processes.
A reliable Type B installation requires clean thermoelements, appropriate process atmosphere, high-purity ceramic protection, suitable compensating cable, correct Type B linearization, cold-junction handling and consideration of high-temperature drift.